Daily quality assurance phantom for ultrasound image guided radiation therapy

基本信息

DOI:
10.1120/jacmp.v8i3.2467
发表时间:
2007-01-01
影响因子:
2.1
通讯作者:
Hilts, Michelle
中科院分区:
医学4区
文献类型:
Article
作者: Drever, Laura;Hilts, Michelle研究方向: -- MeSH主题词: --
关键词: --
来源链接:pubmed详情页地址

文献摘要

A simple phantom was designed, constructed, tested, and clinically implemented for daily quality assurance ( QA) of an ultrasound-image-guided radiation therapy ( US-IGRT) system, the Restitu Ultrasound system ( Resonant Medical, Montreal, QC). The phantom consists of a high signal echogenic background gel surrounding a low signal hypoechoic egg-shaped target. Daily QA checks involve ultrasound imaging of the phantom and segmenting of the embedded target using the automated tools available on the US-IGRT system. This process serves to confirm system hardware and software functions and, in particular, accurate determination of the target position. Experiments were conducted to test the stability of the phantom at room temperature, its tissue-mimicking properties, the reproducibility of target position measurements, and the usefulness of the phantom as a daily QA device. The phantom proved stable at room temperature, exhibited no evidence of bacterial or fungal invasion in 9 months, and showed limited desiccation ( resulting in a monthly reduction in ultrasound-measured volume of approximately 0.2 cm(3)). Furthermore, the phantom was shown to be nearly tissue-mimicking, with speed of sound in the phantom estimated to be 0.8% higher than that assumed by the scanner calibration. The phantom performs well in a clinical setting, owing to its light weight and ease of operation. It provides reproducible measures of target position even with multiple users. At our center, the phantom is being used for daily QA of the US-IGRT system with clinically acceptable tolerances of +/- 1 cm(3) on target volume and +/- 2 mm on target position. For routine daily QA, this phantom is a good alternative to the manufacturer-supplied calibration phantom, and we recommended that that larger phantom be reserved for less frequent, more detailed QA checks and system calibration.
设计、构建、测试了一种简易体模,并将其应用于临床,用于超声图像引导放射治疗(US - IGRT)系统(瑞思图超声系统,Resonant Medical公司,加拿大魁北克省蒙特利尔市)的日常质量保证(QA)。该体模由高信号回声背景凝胶组成,凝胶内包围着一个低信号弱回声的蛋形靶区。日常QA检查包括对体模进行超声成像,并使用US - IGRT系统上的自动化工具对嵌入的靶区进行分割。这一过程用于确认系统硬件和软件功能,特别是对靶区位置的准确测定。进行了实验以测试体模在室温下的稳定性、其仿组织特性、靶区位置测量的可重复性以及该体模作为日常QA设备的实用性。体模在室温下稳定,在9个月内未出现细菌或真菌侵入的迹象,并且干燥程度有限(导致超声测量体积每月减少约0.2立方厘米)。此外,体模被证明几乎具有仿组织特性,体模中的声速估计比扫描仪校准所采用的声速高0.8%。由于其重量轻且操作简便,该体模在临床环境中表现良好。即使由多个使用者操作,它也能提供可重复的靶区位置测量结果。在我们中心,该体模正用于US - IGRT系统的日常QA,靶区体积的临床可接受公差为±1立方厘米,靶区位置为±2毫米。对于常规的日常QA,这种体模是制造商提供的校准体模的良好替代品,我们建议将较大的体模保留用于不太频繁但更详细的QA检查和系统校准。
参考文献(9)
被引文献(0)

数据更新时间:{{ references.updateTime }}

Hilts, Michelle
通讯地址:
--
所属机构:
--
电子邮件地址:
--
免责声明免责声明
1、猫眼课题宝专注于为科研工作者提供省时、高效的文献资源检索和预览服务;
2、网站中的文献信息均来自公开、合规、透明的互联网文献查询网站,可以通过页面中的“来源链接”跳转数据网站。
3、在猫眼课题宝点击“求助全文”按钮,发布文献应助需求时求助者需要支付50喵币作为应助成功后的答谢给应助者,发送到用助者账户中。若文献求助失败支付的50喵币将退还至求助者账户中。所支付的喵币仅作为答谢,而不是作为文献的“购买”费用,平台也不从中收取任何费用,
4、特别提醒用户通过求助获得的文献原文仅用户个人学习使用,不得用于商业用途,否则一切风险由用户本人承担;
5、本平台尊重知识产权,如果权利所有者认为平台内容侵犯了其合法权益,可以通过本平台提供的版权投诉渠道提出投诉。一经核实,我们将立即采取措施删除/下架/断链等措施。
我已知晓