喵ID:L5qZcL免责声明

Influence of conformity on the wear of total knee replacement: An experimental study.

基本信息

DOI:
10.1177/0954411917746433
发表时间:
2018-03
期刊:
Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine
影响因子:
--
通讯作者:
Jennings LM
中科院分区:
其他
文献类型:
Journal Article
作者: Brockett CL;Carbone S;Fisher J;Jennings LM研究方向: -- MeSH主题词: --
关键词: --
来源链接:pubmed详情页地址

文献摘要

Wear of total knee replacement continues to be a significant factor influencing the clinical longevity of implants. Historically, failure due to delamination and fatigue directed design towards more conforming inserts to reduce contact stress. As new generations of more oxidatively stable polyethylene have been developed, more flexibility in bearing design has been introduced. The aim of this study was to investigate the effect of insert conformity on the wear performance of a fixed bearing total knee replacement through experimental simulation. Two geometries of insert were studied under standard gait conditions. There was a significant reduction in wear with reducing implant conformity. This study has demonstrated that bearing conformity has a significant impact on the wear performance of a fixed bearing total knee replacement, providing opportunities to improve clinical performance through enhanced material and design selection.
全膝关节置换的磨损仍然是影响植入物临床使用寿命的一个重要因素。从历史上看,由于分层和疲劳导致的失效促使设计朝着更贴合的衬垫方向发展,以降低接触应力。随着新一代更具氧化稳定性的聚乙烯的研发,在关节面设计上有了更多的灵活性。本研究的目的是通过实验模拟来研究衬垫贴合度对固定承重全膝关节置换磨损性能的影响。在标准步态条件下对两种几何形状的衬垫进行了研究。随着植入物贴合度的降低,磨损显著减少。这项研究表明,关节面贴合度对固定承重全膝关节置换的磨损性能有重大影响,为通过优化材料和设计选择来提高临床性能提供了机会。
参考文献(0)
被引文献(0)
Long-Term Survivorship and Clinical Outcomes Following Total Knee Arthroplasty
DOI:
10.1016/j.arth.2015.05.052
发表时间:
2015-12-01
期刊:
JOURNAL OF ARTHROPLASTY
影响因子:
3.5
作者:
Jauregui, Julio J.;Cherian, Jeffrey J.;Mont, Michael A.
通讯作者:
Mont, Michael A.
Wear of moderately cross-linked polyethylene in fixed-bearing total knee replacements
DOI:
10.1177/0954411912445265
发表时间:
2012-01-01
期刊:
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE
影响因子:
1.8
作者:
Brockett, Claire L.;Jennings, Louise M.;Fisher, John
通讯作者:
Fisher, John
Lower prosthesis-specific 10-year revision rate with crosslinked than with non-crosslinked polyethylene in primary total knee arthroplasty.
DOI:
10.3109/17453674.2015.1065046
发表时间:
2015
期刊:
Acta orthopaedica
影响因子:
3.7
作者:
de Steiger RN;Muratoglu O;Lorimer M;Cuthbert AR;Graves SE
通讯作者:
Graves SE
Relative contributions of design, alignment, and loading variability in knee replacement mechanics
DOI:
10.1002/jor.22169
发表时间:
2012-12-01
期刊:
JOURNAL OF ORTHOPAEDIC RESEARCH
影响因子:
2.8
作者:
Fitzpatrick, Clare K.;Clary, Chadd W.;Rullkoetter, Paul J.
通讯作者:
Rullkoetter, Paul J.
Long-Term Clinical Outcomes and Survivorship of Press-Fit Condylar Sigma Fixed-Bearing and Mobile-Bearing Total Knee Prostheses in the Same Patients
DOI:
10.2106/jbjs.m.01130
发表时间:
2014-10-01
期刊:
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME
影响因子:
5.3
作者:
Kim, Young-Hoo;Park, Jang-Won;Kim, Yoon-Hong
通讯作者:
Kim, Yoon-Hong

数据更新时间:{{ references.updateTime }}

关联基金

Experimental Equipment Call - University of Leeds
批准号:
EP/M028143/1
批准年份:
2015
资助金额:
469.65
项目类别:
Research Grant
Jennings LM
通讯地址:
--
所属机构:
--
电子邮件地址:
--
免责声明免责声明
1、猫眼课题宝专注于为科研工作者提供省时、高效的文献资源检索和预览服务;
2、网站中的文献信息均来自公开、合规、透明的互联网文献查询网站,可以通过页面中的“来源链接”跳转数据网站。
3、在猫眼课题宝点击“求助全文”按钮,发布文献应助需求时求助者需要支付50喵币作为应助成功后的答谢给应助者,发送到用助者账户中。若文献求助失败支付的50喵币将退还至求助者账户中。所支付的喵币仅作为答谢,而不是作为文献的“购买”费用,平台也不从中收取任何费用,
4、特别提醒用户通过求助获得的文献原文仅用户个人学习使用,不得用于商业用途,否则一切风险由用户本人承担;
5、本平台尊重知识产权,如果权利所有者认为平台内容侵犯了其合法权益,可以通过本平台提供的版权投诉渠道提出投诉。一经核实,我们将立即采取措施删除/下架/断链等措施。
我已知晓